DJZ-805 Electrolyte Analyzer applies
advanced ISE (Ion Selection Electrode)
technology to directly and rapidly measure
the concentrations of K, Na, Cl and Ca ions,
as well as pH values in whole blood, serum,
plasma and urine
solutions (diluted). It is an accurate and
reliable clinical testing instrument with
high- degree automation, complete functions
and easy operations.
Technical Specifications
|
Measurement Item
|
Unit
|
Range of
Measurement
|
Accuracy
|
Linearity
|
Repeatability
(CV)
|
|
K+
|
mmol/L
|
1.500
~
10.000
|
0.001
|
0.3
|
1%
|
|
Na+
|
mmol/L
|
50.0
~
200.0
|
0.1
|
3.0
|
1%
|
|
Cl-
|
mmol/L
|
50.0
~
200.0
|
0.1
|
3.0
|
1%
|
|
Ca++
|
mmol/L
|
0.500
~
3.500
|
0.001
|
0.2
|
2%
|
|
pH
|
|
6.400
~
8.400
|
0.001
|
--
|
1%
|
Packing Details
:
490mm×410mm×430mm
N.W.16kgs
G.W.19kgs
Reagents :
1. Calibration
solution
500ml×6
2. Slop
solution
500ml
3. Saturated
KCLsolution
10ml
4.
Protein-removing solution
5ml
5. Electrode
solution 10ml
Features:
The
instrument automatically performs two point
calibration and measurement. The results are
directly displayed on the LCD, and printed
out by the built-in printer.
The
instrument applies ISE (Ion Selection
Electrode), which gives the instrument the
advantages of minimum maintenance, easy
assembly and disassembly, as well as stable
and reliable measurements.
The
instrument applies separate gas/liquid
back-flushing technique, resulting in more
thorough and clean washing.
Automatic
troubleshooting: The instrument
automatically performs solution checking
after it is switched on, and diagnoses any
troubles during the two-point calibration
process. The instrument is also equipped
with service sequences to examine the
operating status of the instrument one by
one, and show the conditions of the
electrodes.
Automatic
storage of measurement results: The
instrument can automatically save up to 400
measurement results, and print out the
results according to the patient
identification number.
The
instrument is equipped with a RS232 serial
port, via which the instrument transfers its
stored data to the connected computer
system.
|